Hose barbed connection Messturbinen Schlauchanschluss-VCT

Turbine flow meter hose barbed connection VCT-HB

Calibrated high level output  (TTL)
Calibrated high level output (TTL)
Temperature probe media exposed
Temperature probe media exposed
Identification memory
Identification memory

Functional description

New designed VCT rotor blades are are running on a high-performance hybrid ball bearing. As a result the VCT turbine meters provides low pressure drop, large turn down ratio an excellent repeatability. An itegrated media exposed temperature probe take readings even on dynamic temperature changes. Turbine and temperature probe defaults are stored in a memory. Rotor speed is dedected without drag and converted to a high level TTL pulse output. Together with a smart VCA flow computer the whole system will configure itself automatically and provide analoge output for both flow rate and media temperature. Output signals are linearized and compensated for fluid viscosity changes.

Product details

Hose connection Ø9,8 to 32mm
Measurement range0,1 to 350l/min
Repeatability± 0,025 % of reading
Temperature probe media exposedType T thermocouple

Temperature Range

Medium-40 to +150°C
Environment-40 to +125°C

Output signal

Turbine frequencymax. 2 kHz (TTL)

Dimensions

Dimension Hose Barbed Connection Messturbinen Mit Schlauchanschluss Vct
TypeInner ØMeasurement rangeHose connectiono Hose connectiono Hose connectiono Hose connectionSWFilter
min.max.DEFMass
VCT[mm][Lit./Min.][Lit./Min.][mm][mm][mm][g][mm][µm]
00056,80,0559,95629,9852110
00106,80,1109,85629,9852110
00207,90,22012,05631,61002110
00309,90,33013,05633,01052110
006012,30,66015,46935,51302110
010015,51,010018,38039,01402110
025021,42,025025,09645,02102720
035027,43,035032,010848,52603220